Article ID Journal Published Year Pages File Type
3117497 American Journal of Orthodontics and Dentofacial Orthopedics 2011 6 Pages PDF
Abstract

IntroductionIn this study, we examined the influence of field of view (FOV) and voxel size on the diagnostic efficacy of cone-beam computed tomography (CBCT) scans to detect erosions in the temporomandibular joint (TMJ).MethodsThe sample consisted of 16 TMJs containing natural or artificially created erosions and 16 normal TMJs. CBCT scans were obtained with 3 imaging protocols differing in the FOV and the size of the reconstructed voxels. Two oral and maxillofacial radiologists scored the scans for the presence or absence of erosions. Diagnostic efficacies of the 3 imaging protocols were compared by using receiver operating curve analysis. For each TMJ imaging protocol, we used thermoluminescent dosimetry chips to measure the absorbed dose at specific organ and tissue sites. Effective doses for each examination were calculated.ResultsAreas under the receiver operating characteristic curves were 0.77 ± 0.05 for the 6-in FOV, 0.70 ± 0.08 for the 9-in FOV, and 0.66 ± 0.05 for the 12-in FOV. The diagnostic efficacy of the 6-in FOV, determined by the area under the curve, was significantly higher than that of the 12-in FOV (P ≤0.05). Effective doses for bilateral TMJ evaluation were 558 μSv for the 6-in FOV, 548 μSv for the 9-in FOV, and 916 μSv for the 12-in FOV.ConclusionsThe diagnostic efficacy of CBCT scans for the evaluation of erosive changes in the TMJ is highest for the 6-in FOV and lowest for the 12-in FOV.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine
Authors
, , , , ,