Article ID Journal Published Year Pages File Type
3117848 American Journal of Orthodontics and Dentofacial Orthopedics 2010 7 Pages PDF
Abstract

IntroductionIn this study, we aimed to determine the effect of bone mineral density (BMD), cortical bone thickness (CBT), screw position, and screw design on the stability of miniscrews.MethodsNinety-six miniscrews of both cylindrical and tapered types were placed in 6 beagle dogs. The BMD and CBT were measured by computerized tomography and correlated with the placement and removal torque and mobility. A regression equation to predict the placement torque was calculated based on BMD, CBT, screw type, and screw position.ResultsThe placement torque showed a positive correlation in the order of removal torque (0.66), BMD of the cortical bone (0.58), and CBT (0.48). Placement and removal torque values were significantly higher in the mandible compared with the maxilla. Tapered miniscrews had higher placement torque than did the cylindrical type (P <0.001). However, the removal torque was similar in both groups. Placement torque was affected by screw position, screw type, and BMD of cortical bone, in that order.ConclusionsBMD of cortical bone, screw type, and screw position significantly influence the primary stability of miniscrews.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine