Abstract

Introduction: The purpose of this in-vitro study was to compare the forces generated by new nonconventional elastomeric ligatures (NCEL) and conventional elastomeric ligatures (CEL) during leveling and aligning phases. Methods: The testing model consisted of 5 stainless steel 0.022-in preadjusted brackets for second premolar, first premolar, canine, lateral incisor, and central incisor. The canine bracket was welded to a sliding bar that allowed for different vertical positions. The forces generated by 3 sizes of wires (0.012-, 0.014-, and 0.016-in superelastic nickel-titanium) with the 2 types of elastomeric ligatures at different amounts of upward canine misalignment (1.5, 3, 4.5, and 6 mm) were recorded. Results and Conclusions: Significant differences between CEL and NCEL were found for all tested variables (P <.01) with the exception of the 0.014- and 0.016-in wires at canine misalignment of 1.5 mm. A noticeable amount of force was generated with the NCEL at all 4 canine positions with all 3 wire sizes (from about 50 to about 150 g). With 4.5 mm of canine misalignment or more, the average amount of released force with the CEL was approximately zero.