Abstract

We aimed to assess the stability over time of dental implants placed in grafted bone in the maxilla using resonance frequency analysis, and to compare the stability of implants placed in grafted and non-grafted bone. Data were collected from 23 patients (15 test and 8 controls) in whom 64 implants (Brånemark system, Nobel Biocare, Göteborg, Sweden) were placed in accordance with the two-stage surgical protocol. In the test group 36 fixtures were placed in grafted bone, and in the control group 28 fixtures were placed in non-grafted bone. Resonance frequency analysis was used to assess the test sites at implant placement and abutment connection. The mean (SD) implant stability quotient (ISQ) for test sites at the time of implant placement was 61.91 (6.68), indicating excellent primary stability, and was 63.53 (5.76) at abutment connection. ISQ values at abutment connection were similar for test and control sites. Implants placed in grafted bone compared favourably with those in non-grafted bone, and showed excellent stability.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine