Article ID Journal Published Year Pages File Type
3131313 Future Dental Journal 2015 5 Pages PDF
Abstract

ObjectivesThis research was carried out to evaluate the bacterial leakage of two different internal implant abutment connections in vitro.Materials and MethodsTwenty dental implants divided into two equal groups were compared; Group 1 fixtures with an internal hexagonal geometry; Group 2 fixtures with a tri-lobe internal connection. A bacterial suspension of Staphylococcus aureus was prepared to obtain a density of 0.5 McFarland standards. All implant abutment assemblies were submerged in sterile tubes containing 4 mL of S. aureus broth culture and were incubated at 37 °C for 14 days. The specimens were disassembled and the inner surfaces of the implants were sampled by sterile paper points. Then the paper points were immersed in test tubes containing sterile BHI broth. From the broth, culture was done on blood agar plates and incubated at 37 °C for 24 h. The resulting colonies were identified by Gram's stain and biochemical reactions.ResultsInternal hexagon implants showed statistically significant higher mean Log10 CFU than Tri-lobe implants.ConclusionBacterial leakage seems to be inevitable but fixture abutment interface geometry plays an important role in the amount of leakage.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine
Authors
, ,