Article ID Journal Published Year Pages File Type
3146940 Journal of Endodontics 2014 4 Pages PDF
Abstract

IntroductionElectronic working length measurement during root canal treatment in teeth with enlarged apical foramina is a challenge. The aims of this in vitro study were (1) to assess the influence of foramen widening on the accuracy of 5 different electronic foramen locators (EFLs) and (2) to compare the accuracy of EFLs in different foramen sizes.MethodsThe following EFLs were used: MiniApex, Root ZXII, iPex, Propex II, and Elements Apex Locator. Each EFL was used in 3 groups (n = 20) of extracted teeth, with foramen diameters of 0.27 mm (G27), 0.47 mm (G47), and 0.72 mm (G72). Working length was measured according to manufacturer's instructions and compared with visual measurements (control method). Results were classified as accurate (equal or differences ≤ 0.05 mm) or inaccurate (differences > 0.5 mm).ResultsIn G27, all EFLs yielded accurate findings (intragroup reliability; Fisher exact test, P < .05), compared with only MiniApex, Root ZXII, and Elements Apex Locator in G47 and G72. MiniApex, Root ZXII, and Elements Apex Locator were similarly accurate regardless of foramen size. iPex and Propex II were the least accurate among the devices tested, and foramen diameter influenced their accuracy, with greater diameters yielding poorer EFL performance.ConclusionsForamen diameter did not influence the accuracy of MiniApex, Root ZXII, and Elements Apex Locator EFLs. iPex and Propex II showed decreased accuracy as foramen size increased.