Article ID Journal Published Year Pages File Type
3147726 Journal of Endodontics 2009 4 Pages PDF
Abstract

The purpose of this study was to examine three different types of nickel-titanium (NiTi) systems (ProTaper [Dentsply Maillefer, Ballaigues, Switzerland], ProTaper for Hand Use [Dentsply Maillefer], and K3 [SybronEndo, Orange, CA]) that were discarded by 3 endodontic clinics. The instruments were evaluated for defects and factors leading to instrument deformation or fracture. A total of 1682 instruments were collected over 16 months and were examined. The location of the defect, if any, was recorded. The overall prevalence of unwinding defects was 3% and fracture 5%; the rates differed significantly between clinics. For one brand (ProTaper) used at two different clinics, a defect rate (fracture and distortion combined) of 7% (clinic A) vs. 13% (clinic B) for Shaping files (P < 0.05), and about 4% vs. 10% for Finishing files (P < 0.05) was observed. Fragments of broken Shaping file were significantly longer in clinic A than for clinic B (P < 0.05). The lowest defect rate was found for K3 instruments: unwinding 1%, and fracture 3%. It was concluded that the defect rates of NiTi instruments were influenced by such factors as the operator, preparation technique and instrument design.

Related Topics
Health Sciences Medicine and Dentistry Dentistry, Oral Surgery and Medicine
Authors
, , , ,