Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3155249 | Journal of Oral and Maxillofacial Surgery | 2007 | 7 Pages |
Abstract
Unilock plates fixed with locking screws have a higher long-term stability than the Reconstruction 2.4 system. A 1-mm gap between the plate and mandible does not lead to early screw failure in the Unilock 2.4 system with locking screws. The Reconstruction 2.7 system seems superior if well contoured, because 2 of those plates reached the runout limit; however, this system is not as easy to handle as the 2.4 systems, and good contouring is difficult to achieve. Therefore, we consider the Unilock 2.4 system with locking screws the best choice.
Related Topics
Health Sciences
Medicine and Dentistry
Dentistry, Oral Surgery and Medicine
Authors
Wiebke MD, DDS, Michael Arzdorf, Berend Linke, Ralf MD, DDS,