Abstract

ObjectiveThe aim of the trial was to compare the time parameters for intubation with the use of the Macintosh (MAC) laryngoscope and Pentax AWS-S100 videolaryngoscope (AWS; Pentax Corporation, Tokyo, Japan) with and without chest compression (CC) by paramedics during simulated cardiopulmonary resuscitation in a cadaver model.MethodsThis was a randomized crossover cadaver trial. Thirty-five paramedics with no experience in videolaryngoscopy participated in the study. They performed intubation in two emergency scenarios: scenario A, normal airway without CC; scenario B, normal airway with continuous CC.ResultsThe median time to first ventilation with the use of the AWS and the MAC was similar in scenario A: 25 (IQR, 22-27) seconds vs. 24 (IQR, 22.5-26) seconds (P = .072). A statistically significant difference in TTFV between AWS and MAC was noticed in scenario B (P = .011). In scenario A, the first endotracheal intubation (ETI) attempt success rate was achieved in 97.1% with AWS compared with 94.3% with MAC (P = .43). In scenario B, the success rate after the first ETI attempt with the use of the different intubation methods varied and amounted to 88.6% vs. 77.1% for AWS and MAC, respectively (P = .002).ConclusionsThe Pentax AWS offered a superior glottic view as compared with the MAC laryngoscope, which was associated with a higher intubation rate and a shorter intubation time during an uninterrupted CC scenario. However, in the scenario without CC, the results for AWS and MAC were comparable.

Related Topics
Health Sciences Medicine and Dentistry Emergency Medicine