Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3251688 | Journal of Orthopaedics | 2015 | 5 Pages |
Abstract
PurposeThis pilot study tested whether FiberWire provides similar protection to steel wire against repair displacement in patella fractures.MethodsThirteen cadaver knees were cyclically loaded with 10 cycles (0–90° flexion) and fracture displacement was recorded. Fixation methods were also tested in load to failure (>3 mm displacement).ResultsThere was no difference between wire types in fracture displacement (1.4 mm ± 0.33 mm vs 1.2 mm ± 0.34 mm, respectively; p = 0.418) or in the load to failure (714.7 N ± 110.9 N vs 744.5 N ± 92.8 N, respectively; p = 0.360).ConclusionFiberWire provides similar protection to steel wire against repair displacement after fixation of patellar fractures.
Related Topics
Health Sciences
Medicine and Dentistry
Emergency Medicine
Authors
Tony L. Bryant, Christopher L. Anderson, Christopher G. Stevens, Bryan P. Conrad, Heather K. Vincent, Kalia K. Sadasivan,