Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3271358 | Journal of Clinical Densitometry | 2008 | 6 Pages |
Abstract
Researchers conducting longitudinal studies face challenges as technology evolves and data collection equipment is upgraded from older to newer models. In this study, we report precision error and associations from correlation, regression equations, and Bland-Altman plots that compare results of distal radius images from the XCT 2000 and XCT 3000 peripheral quantitative computed tomography densitometers at the 4% and 20% distal radius site. The mean difference expressed as a percent of the measurement's mean was approximately 5% for cortical thickness, periosteal and endosteal circumferences, and total bone size at the 20% site, and less than 2% for all other measures. There were strong associations between bone parameter measures from the XCT 3000 and XCT 2000 with correlation coefficients ranging from 0.92 to 1.0. Strong associations between the 2 machines allow for the comparison of results with minor adjustments.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Endocrinology, Diabetes and Metabolism
Authors
Cory Koenig, Howard Wey, Teresa Binkley,