Article ID Journal Published Year Pages File Type
3271689 Journal of Clinical Densitometry 2008 4 Pages PDF
Abstract
In this study, the short-term reproducibility and agreement between individual units of dual X-ray and laser (DXL) Calscan (version 2), a new bone densitometry device based on dual X-ray absorptiometry and Laser technique, were evaluated. The variations in bone mineral density (BMD) and T-score between the right and the left foot were also studied retrospectively in 334 individuals. The short-term CV%, based on 19 healthy individuals scanned twice on 4 different units, was 2.1%, and the average CV% per scanner ranged from 1.1% to 1.8%. The short-term CV% in 33 elderly individuals (aged 83 ± 6 yr) was 3.9% and in a subgroup with a calcaneal T-score of −2.5 or below (n = 16) 5.1%. There was no statistical difference between the average BMD of the right and left foot, and (r2) between the right and left side was 0.84. The observed short- and long-term reproducibilities, as well as the acceptable variation between individual machines, show that this version of Demetech DXL Calscan is suitable for measuring BMD.
Related Topics
Health Sciences Medicine and Dentistry Endocrinology, Diabetes and Metabolism
Authors
, ,