Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3271708 | Journal of Clinical Densitometry | 2007 | 6 Pages |
Abstract
The purpose of this study was to carry out an in vivo cross calibration for forearm bone mineral density (BMD) between a Hologic QDR2000 (Hologic Inc., Bedford MA, USA) and Lunar Prodigy (GE Healthcare, Madison WI, USA) during equipment upgrade. Nineteen subjects (17 females and 2 males, mean age 57Â yr, range: 42-79Â yr) attending for routine dual energy X-ray absorptiometry scanning were recruited. BMD of the nondominant forearm was measured on both instruments. Cross-calibration equations were derived for total forearm, ultradistal radius and ulna, and 33% radius and ulna. A Bland & Altman plot was used to calculate the mean difference and limits of agreement between instruments. There were significant differences in BMD at all sites. The Prodigy BMD was 15% higher at the total forearm and 20-25% higher in the cortical regions of the 33% ulna and 33% radius. The differences are smaller in the ultradistal regions, as it appears that the Prodigy underestimates BMD at low BMD. The standard error of estimate about the cross calibration was such that it cannot be used to transfer individual patients between instruments, but could be applied to clinical trial data.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Endocrinology, Diabetes and Metabolism
Authors
D. Pearson, B. Horton, D.J. Green,