Article ID Journal Published Year Pages File Type
3303747 Gastrointestinal Endoscopy 2013 4 Pages PDF
Abstract

BackgroundZenker's diverticulum (ZD) is a rare disorder but is associated with significant morbidity. Cricopharyngeal (CP) myotomy is the mainstay of treatment, and various flexible endoscopic techniques have been used for division of the septum. However, there is a constant need for improvement in accessories.ObjectiveTo evaluate the safety and effectiveness of a new electrocautery endoscopic scissor for CP myotomy in patients with symptomatic ZD.DesignObservational human study.SettingTertiary-care hospital.PatientsThis study involved 3 patients with symptomatic ZD.InterventionFlexible endoscopic CP myotomy was performed by using a novel scissors-type grasping device. CP myotomy involved 4 steps: (1) opening of the forceps, (2) grasping the muscle fiber, (3) closure of the forceps with application of gentle traction, and (4) dissection of muscle fibers by using cutting current. Intraprocedural bleeding was controlled with the same instrument by grasping vessels and applying coagulation current.Main Outcome MeasurementsOverall feasibility and performance, procedure time to achieve complete CP myotomy.ResultsCP myotomy was successfully performed in all patients. Mean procedure time was 10.6 minutes. There were no major adverse events. Minor intraprocedure bleeding occurred in 1 patient.LimitationsSingle arm, limited number of patients.ConclusionThe new instrument has potential advantages in comparison with standard instruments used for CP myotomy. The advantages of this new technique are better control of cutting and hemostatic abilities.

Related Topics
Health Sciences Medicine and Dentistry Gastroenterology
Authors
, ,