Article ID Journal Published Year Pages File Type
3348151 Diagnostic Microbiology and Infectious Disease 2007 6 Pages PDF
Abstract

We compared an in-house Treponema pallidum IgM immunoblot (IB) with a 19S fluorescent treponemal antibody absorption (IgM) test during routine use for the diagnosis of congenital syphilis (CS) in a national reference laboratory in a nonendemic setting. The overall agreement between the assays was high (97%), and 19S positive samples had at least 2 reactive bands in the IB. The high agreement is mainly caused by the large number of negative results (95%). If the 19S is taken as the gold standard, the estimate sensitivity of the IB was at least 88% with a specificity of 97.2%. Analysis of the discrepancies revealed that the IB was positive with 1 or 2 specific bands in 2.8% of the cases, whereas 19S was negative, possibly indicating higher sensitivity of the IB. We conclude that the IB is a sensitive method to detect contact with T. pallidum in neonates and can replace the 19S in routine laboratory screening for CS cases.

Related Topics
Life Sciences Immunology and Microbiology Applied Microbiology and Biotechnology
Authors
, , , ,