Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
3848059 | American Journal of Kidney Diseases | 2014 | 4 Pages |
Abstract
This study recommends that urine samples collected for the assessment of KIM-1 or NGAL are immediately cooled to 4°C and subsequently frozen at â80°C within 2 days. At â80°C, both KIM-1 and NGAL concentrations are stable for at least 6 months.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Nephrology
Authors
Mathijs MSc, Jeroen K. MD, PhD, Johan PhD, Luuk B. MD, PhD,