With the current power morcellators, the amount of tissue spread peaks and is independent of uterine weight after a certain cutoff point (in this study 350 g). There is a relative inefficiency in the rotational mechanism because mostly small tissue strips are created. These small tissue strips occur increasingly later on in the procedure. Because small tissue strips are inherently more prone to scatter by the rotational mechanism of the morcellator, the risk of tissue spread is highest at the end of the morcellation procedure. This means that LSH and laparoscopic hysterectomy procedures may be at higher risk for tissue scatter than total laparoscopic hysterectomy. Finally, engineers should evaluate how to create only large tissue strips or assess alternatives to the rotational mechanism.