3985525 European Journal of Surgical Oncology (EJSO) 2014 5 Pages PDF
Abstract

A prospective study was lead in order to analyze the accuracy of an X-ray device settled in the operating room for margin assessment, when performing breast-conserving surgery.Patients and methodsOne hundred and seventy patients were included. All lesions were visible on the preoperative mammograms. An intraoperative X-ray of the lumpectomy specimen was systematically performed for margins assessment. Final histological data were collected and the accuracy of intraoperative specimen radiography (IOSR) for margin assessment was analyzed.ResultsIOSR allowed an evaluation of margins status in 155 cases (91.2%). After final histological examination, the positive margins rate would have been 6.5% if margin assessment had relied only on IOSR.ConclusionMargin assessment with a two-dimensional X-ray device would have allowed the achievement of negative margins in 93.5% of the cases. Moreover, this procedure allows important time-saving and could have a substantial economical impact.

Related Topics
Health Sciences Medicine and Dentistry Oncology