Article ID Journal Published Year Pages File Type
4003817 American Journal of Ophthalmology 2010 7 Pages PDF
Abstract
The 7 OCT devices presented differing macular thickness measurement values; the lowest value was with Copernicus, and the highest was with Spectralis HRA+OCT (Heidelberg Engineering). The new generation of SD OCT devices has good intraoperator reproducibility, but the Spectralis presents the highest reproducibility together with the best interoperator agreement. The software, and in particular the algorithm, used is the most important factor regarding reproducibility differences in macular thickness measurement. The main result of our study is that macular thickness absolute value differs for each device. For this reason, the devices are not interchangeable.
Related Topics
Health Sciences Medicine and Dentistry Ophthalmology
Authors
, , , ,