Article ID Journal Published Year Pages File Type
4004392 American Journal of Ophthalmology 2009 11 Pages PDF
Abstract
Cirrus OCT measures RT ∼43 μm greater than Stratus OCT, which is likely attributable to Cirrus OCT detection of the outer band of the retinal pigment epithelium vs Stratus OCT detection of the inner/outer segment photoreceptor junction. All zones showed a wide degree of variability in correlation. Both Cirrus cube scans provide similar data. Future studies comparing these systems might consider utilizing average macular thickness values, which reflect macular volume and may provide more consistent measurements.
Related Topics
Health Sciences Medicine and Dentistry Ophthalmology