PurposeTo evaluate the repeatability of different lens densitometry analysis modes performed with Scheimpflug imaging using the Pentacam HR device.SettingUniversity eye hospital, Frankfurt am Main, Germany.DesignCross-sectional study.MethodsEyes were analyzed under mydriasis during 2 examinations performed by the same examiner. Three single measurements were taken to assess lens densitometry via 3 modes as follows: peak (maximum), linear (vertical axis), and 3-dimensional (3-D). Measurements were also taken using 3 cylindrical reference blocks (P) with different diameters (2.0 mm, 4.0 mm, 6.00 mm) composing the whole lens depth. Repeatability was tested using Bland-Altman analysis. As statistical parameters, the coefficient of repeatability (CoR) and the relative repeatability (RR) were applied.ResultsThe study analyzed 105 eyes. The CoR of the peak modus was 5.6 (RR, 29.8%); of the linear modus, 0.9 (RR, 8.8%); and of the 3-D modus, 0.36 (RR, 3.6%). The values of the P mode were lower, with a CoR between 0.19 and 0.30 (RR, 2.1%-3.3%).ConclusionsLens densitometry using Scheimpflug imaging was highly repeatable. However, repeatability was dependent on the analysis mode used. Repeatability decreased with increasing opacification.Financial DisclosureNo author has a financial or proprietary interest in any material or method mentioned.