Article ID Journal Published Year Pages File Type
4019635 Journal of Cataract & Refractive Surgery 2009 4 Pages PDF
Abstract

PurposeTo estimate the rotation stability of the bag-in-the-lens (BIL) intraocular lens (IOL) (Morcher 89A) over time and to assess whether the rotational stability of the IOL design is suitable for toric corrections.SettingDepartment of Ophthalmology, Antwerp University Hospital, Edegem, Belgium.MethodsThis prospective study evaluated patients who had cataract surgery and BIL IOL implantation. Postoperative red reflex photographs were taken 1 day, 5 weeks, and 6 months after surgery. Intraocular lens rotation was obtained by defining a triangle between a scleral blood vessel, the IOL haptic, and the IOL optic center. Changes in the triangle during the follow-up were evaluated using purpose-designed software written in MatLab.ResultsFifty-nine eyes of 49 patients (mean age 68.0 years ± 11.9 [SD]) were included. There was little postoperative IOL rotation (mean 0.05 ± 2.02 degrees) between 1 day and 5 weeks (n = 46 eyes) and between 5 weeks and 6 months (mean 0.36 ± 1.39 degrees) (n = 15 eyes). The IOL rotation remained unchanged from 1 to 6 months postoperatively (P = .327, analysis of variance). There was no correlation between IOL rotation and patient age (r2 = 0.011) or IOL power (r2 = 0.003). Postoperative IOL rotations were not different between left eyes and right eyes (P = .862, t test).ConclusionThe BIL IOL showed good rotation stability, making it suitable for toric correction.

Related Topics
Health Sciences Medicine and Dentistry Ophthalmology
Authors
, , , ,