Article ID Journal Published Year Pages File Type
4020960 Journal of Cataract & Refractive Surgery 2007 5 Pages PDF
Abstract

Endothelial decompensation is a serious complication of phakic intraocular lens (pIOL) implantation and is a major concern during the postoperative period. We report 3 eyes in which the same foldable angle-supported pIOL was implanted to correct high myopia. Rapid and severe postoperative endothelial cell loss occurred in all 3 eyes. An over-sized pIOL that induced excessive vaulting into the anterior chamber was the main risk factor. In 2 eyes, the pIOL was explanted uneventfully; 1 eye required Descemet's stripping automated endothelial keratoplasty because of total endothelial decompensation. These cases illustrate the importance of accurate sizing of foldable angle-supported anterior chamber pIOLs to avoid excessive vaulting. They also highlight the importance of regular follow-up and preventive pIOL explantation as soon as significant endothelial cell loss is detected.

Related Topics
Health Sciences Medicine and Dentistry Ophthalmology
Authors
, , , , , , ,