Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4022200 | Journal of Cataract & Refractive Surgery | 2006 | 6 Pages |
Abstract
Under both illuminance conditions, the Bernell card system with the cobalt filter measured a larger pupil size than the Colvard pupillometer. The measurement differences between the techniques were most pronounced at the lower illumination. The limits of agreement were larger under the lower illumination, indicating more variation between techniques. This study suggests that the Bernell card system with cobalt illumination provides a generous measurement of the pupil size compared with the Colvard pupillometer, which makes it an appropriate and cost-effective screening tool for refractive surgery evaluation.
Related Topics
Health Sciences
Medicine and Dentistry
Ophthalmology
Authors
Elyse L. OD, FAAO, Sabah OD, Louis E. MD,