Article ID Journal Published Year Pages File Type
4047396 Arthroscopy: The Journal of Arthroscopic & Related Surgery 2006 6 Pages PDF
Abstract

PurposeTo compare a standard suture material with a newer material using multiple arthroscopic knot configurations and to evaluate the biomechanical performance of a new sliding-locking knot compared with 4 surgical standards.Type of StudyControlled laboratory study.MethodsFive knots were evaluated (Weston, Tennessee, Duncan, SMC, and the new San Diego knot) using 2 suture materials, No. 2 Ethibond (Ethicon, Somerville, NJ) or No. 2 Fiberwire (Arthrex, Naples, FL). Eight samples were tested for each knot-suture configuration. Samples were pretensioned to 10 N and then loaded from 10 to 45 N for 1,000 cycles. Intact knots were loaded to failure.ResultsFiberwire had significantly higher load-to-failure (276 ± 24 N) compared with Ethibond (111 ± 13 N) (P < .001), although there was no significant difference as a function of knot configuration. Of the 40 Fiberwire knots, 3 failed by early slippage during cyclic loading and 8 slipped at very low tension during load-to-failure. None of the Ethibond knots and none of the San Diego knots failed by early slippage.ConclusionsSurface characteristics and suture construction affect the tendency for knot slippage.Clinical RelevanceSurgeons should understand the impact of handling characteristics, frictional properties, and ultimate failure load when selecting suture materials and knots for arthroscopic repair.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation