Article ID Journal Published Year Pages File Type
4050784 Clinical Biomechanics 2012 5 Pages PDF
Abstract

BackgroundDistal femur fractures continue to be a complex surgical problem for which the incidence is increasing. Presently, there is a need for different constructs to address these complex fractures. This study attempts to define the biomechanical properties of several implants.MethodsA novel, prototype locking retrograde intramedullary nail and the Russell–Taylor femoral retrograde nail were tested at non-destructive, physiological, axial mode load strength using a young, synthetic bone model for a medial segmental shaft defect in the supracondylar region of the distal femur (medial gap of 10 mm, 65 mm proximal to the distal joint and parallel to the knee axis). Each specimen was compressively loaded and unloaded to the peak load for 80,000 cycles at a 0.5 Hz frequency. These were compared to the results from the same lab of the retrograde Trigen intramedullary nail. Motion and peak displacement were measured across the fracture site as a reflection of construct stability.FindingsPrevious testing demonstrated that Trigen intramedullary nail had significantly less motion across the gap and increased overall stiffness of the construct (P < 0.05) compared to both Russell–Taylor and prototype nails.InterpretationLocking technology used in a nail biomechanically appears to lead to more micro-motion across the fracture gap and to less stiffness in this construct. Further research needs to be invested into intramedullary, locking technology before introducing it into clinical practice.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation