Article ID Journal Published Year Pages File Type
4050942 Clinical Biomechanics 2010 9 Pages PDF
Abstract

BackgroundShort uncemented metaphyseally anchored femoral endoprostheses are becoming popular and are proposed to be less invasive than longer conventional implants. However, it is proposed here that shortening femoral endoprostheses can increase the risk of periprosthestic fracture.MethodsA simple analytical model of a femoral hip implant was developed to estimate the risk of bone overload for varying implant size, implantation geometry, implantation force and bone quality. The load capacity of a particular short implant design in poor quality cadaveric bone specimens was also measured experimentally, to validate the model.FindingsThe model demonstrated a high risk of bone overload for a short endoprosthesis in poor quality bone. The experimental results and a clinical example of failure, to which the model was applied, supported this finding. Bone stresses increased with decreasing implant length and diameter, varus implantation, incomplete seating and high implantation forces, approaching the strength of good quality bone in extreme cases.InterpretationCorrect implantation and patient selection is essential for short femoral endoprostheses.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , ,