Article ID Journal Published Year Pages File Type
4060122 The Journal of Arthroplasty 2015 5 Pages PDF
Abstract

PSI software adjusts preoperative planning to accommodate differences in implant design. Such adjustments may influence the accuracy of intraoperative jig placement, bone resection, or component placement. Our purpose was to determine whether implant design influences PSI accuracy. 96 and 123 PSI TKA were performed by a single surgeon using two different implant systems and identical PSI software. Femoral coronal alignment outliers were greater for Implant 1 (23.9% Implant 1 vs. 13.4% Implant 2; P = 0.050). Tibial coronal alignment outliers were greater for Implant 2 (10.9% Implant 1 vs. 22.7% Implant 2; P = 0.025). There was no difference in overall mechanical axes. Differences in implant design can influence bone resection and component alignment. PSI software rationale must align with surgeons’ intraoperative goals.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,