Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4060387 | The Journal of Arthroplasty | 2015 | 6 Pages |
Abstract
Robotic computerized instrumentation that guides bone preparation and cup implantation in total hip arthroplasty was studied. In 38 patients (43 hips) intraoperative cup inclination and anteversion were validated by postoperative CT scans. Planned inclination was 39.9° ± 0.8° and with robotic instrumentation was 38. 0° ± 1.6° with no outliers of 5°; on the postoperative CT scan there were 5 outliers (12%). Planned anteversion was 21.2° ± 2.4° and intraoperatively was 20.7° ± 2.4° with no outlier of 5°; on the CT there were 7 outliers (16%). The center of rotation (COR) was superior by a mean 0.9 ± 4.2 mm and medial by 2.7 ± 2.9 mm. This robotic instrumentation achieved precision of inclination in 88%, anteversion in 84% and COR in 81.5%.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Vaibhav Kanawade, Lawrence D. Dorr, Scott A. Banks, Zenan Zhang, Zhinian Wan,