Article ID Journal Published Year Pages File Type
4061613 The Journal of Arthroplasty 2010 4 Pages PDF
Abstract

Increasing latex hypersensitivity among patients and health care workers has prompted the development of latex-free surgical gloves. Latex-free gloves must perform equally as the existing latex standard. We analyzed perforation rates in a clinical trial comparing latex and a latex-free alternative during primary hip and knee arthroplasty. The overall latex glove perforation rate was 8.4% compared with 21.6% for the latex-free alternative (χ2P < .001). The operation perforation rate for latex gloves was 34.4% compared with 80% for latex-free gloves (χ2P < .001). We suggest that the latex-free glove tested cannot provide a reliable barrier between the surgeon and the patient. As such, we question the safety of these gloves and the standards sets by the regulators.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,