Article ID Journal Published Year Pages File Type
4061882 The Journal of Arthroplasty 2008 5 Pages PDF
Abstract

We reviewed 40 Corin Cormet 2000 (Corin, Cirencester, UK) metal-on-metal resurfacing hips, in 36 patients, for the presence of femoral neck narrowing. A neck-to-prosthesis ratio was calculated by dividing the diameter of the femoral neck with that of the implant. This ratio was measured on plain anteroposterior pelvis radiographs taken immediately and 2 years postoperation. Subsequent radiographs were measured up to a maximum 7 years (mean, 5.3 years) postoperation. Femoral neck narrowing was observed in 90% of hips at 2 years, with the average neck narrowing ranging from a ratio of 0.865 to 0.811. Importantly, no further narrowing occurred beyond this point up to 7 years postoperation. We described a simple reproducible method of measuring neck narrowing on plain radiographs and discuss possible causal factors for neck narrowing after hip resurfacing.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , ,