Article ID Journal Published Year Pages File Type
4061939 The Journal of Arthroplasty 2012 6 Pages PDF
Abstract

To evaluate the effects of distal design of a proximally coated femoral component on periprosthetic bone remodeling, we prospectively performed 21 one-stage bilateral total hip arthroplasties using a distally tapered and a distally cylindrical stem with the same proximal design, randomized to side. All hips showed good outcomes clinically and radiographically at the final follow-up, average of 7 years postoperatively. Cancellous condensation was always found in Gruen's zones 2 and 6 around the cylindrical stem and in regions between zones 2 and 3 and between zones 6 and 5 around the tapered stem. Bone mineral density of Gruen's zones 2 and 6 was significantly lower around the tapered stem. These results suggested more distal loading in hips with the tapered stem than in those with the cylindrical stem.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , , ,