Article ID Journal Published Year Pages File Type
4062280 The Journal of Arthroplasty 2010 6 Pages PDF
Abstract

To evaluate the effectiveness of the Bipolar Sealer 6.0-VT (BPS 6.0-VT) (Tissue Link Medical, Inc, Dover, NH) in reducing blood loss compared with a conventional electrocautery (Erbe ICC 350) (Erbe ICC 350 (ERBE Elektromedizin GmbH, Tübingen, Germany). A prospective randomized trial of 105 patients with primary total hip arthroplasty (Erbe ICC 350: 50 patients; BPS 6.0-VT: 55 patients). On the basis of the calculation of the preoperative blood volume the total and postoperative blood loss were calculated. No statistically significant difference could be shown for total intraoperative and postoperative blood loss (group A 1846 mL; group B 1740 mL) (t test). Due to the lack of reduction of blood loss and the high costs of the BPS 6.0-VT, its use is not recommended in primary total hip arthroplasty.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,