Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4063050 | The Journal of Arthroplasty | 2009 | 6 Pages |
Abstract
Valgus alignment of femoral resurfacing components has been advocated to reduce proximal femur loading and thus minimize the risk for femoral neck fractures. However, such reduction in loading may exacerbate undesirable stress shielding. This study examined the effect of extreme implant orientations (±15°) and stem canal overreaming on initial bone remodeling stimulus using finite element models. The changes in implant-cement interface stresses due to implant alignment were also evaluated. The valgus model showed increased initial bone resorption stimulus, which extended distally and peripherally around the femoral neck. The peak implant-cement interface shear stress for the varus model was 10.9 MPa, exceeding the interface shear strength. Overreaming of the stem canal eliminated distal tip loading, but proximal stress shielding was still unavoidable. These data show bone loading and interface fixation trends emanating from valgus and varus implant positions that will be of interest to practicing physicians.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Kevin L. PhD, Judd S. PhD, Steven M. PhD, Richard E. PhD, FRCS, FRCS(Orth), Michael T. PhD,