Article ID Journal Published Year Pages File Type
4063117 The Journal of Arthroplasty 2006 8 Pages PDF
Abstract

An instrumented tibial prosthesis was developed to measure forces in vivo after total tibial arthroplasty. This prosthesis was implanted in a 67-kg, 80-year-old man. The prosthesis measured forces at the 4 quadrants of the tibial tray. Tibial forces were measured postoperatively during rehabilitation, rising from a chair, standing, walking, and climbing stairs. By the sixth postoperative week, the peak tibial forces during walking averaged 2.2 times body weight (BW). Stair climbing increased from 1.9 times BW on day 6 to 2.5 times BW at 6 weeks. This represents the first direct in vivo measurement of tibial forces, which should lead to refined surgical techniques and enhanced prosthetic designs. Technical design improvements will enhance function, quality of life, and longevity of total knee arthroplasty.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , ,