Article ID Journal Published Year Pages File Type
4063207 The Journal of Arthroplasty 2011 4 Pages PDF
Abstract
Four years after primary implantation of an ESKA Cut short-stemmed prosthesis, a 61-year-old patient suffered fracture of the implant. The combination of missing proximal support of the prosthesis as well as enormous strain on the distal third of the stem can be considered as main reason for the fatigue failure. The special design of the modular implant with a small diameter and a spongy metal surface can be regarded as a contributing factor. This unique case demonstrates a possible failure mechanism of short- and ultra short-stemmed prostheses.
Keywords
Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , ,