Article ID Journal Published Year Pages File Type
4063567 The Journal of Arthroplasty 2008 7 Pages PDF
Abstract

We present, for the first time, a detailed damage assessment of a large-diameter metal-on-metal (L-MOM) hip prosthesis to show the extent of surface damage that can occur in a patient after multiple dislocations. The patient was a man (51 years old) who dislocated 8 times and was finally revised at 27 months. Radiographically, the cup was malpositioned with 65° lateral opening and 15° retroversion. The retrieved cup was a 1-piece, 38-mm Co-Cr-Mo (M2a; Biomet, Warsaw, Ind) with a titanium-alloy backing. The retrieved components demonstrated all known modes of wear, including a polished wear scar, multidirectional scratching, “stripe” wear, surface contamination of titanium-alloy, front face wear, and backside wear. The clinical significance is that cup positioning remains critical regardless of whether a large diameter head is used or not.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , ,