Article ID Journal Published Year Pages File Type
4063619 The Journal of Arthroplasty 2009 6 Pages PDF
Abstract

This prospective, randomized protocol evaluated femoral head penetration after total hip arthroplasty in a young population. Forty-five patients randomly received either a cross-linked or conventional ultrahigh-molecular-weight polyethylene (UHMWPE) liner in a noncemented hemispheric cup (Trilogy, Zimmer, Warsaw, Ind) with a 28-mm femoral head. Radiostereometric analysis film pairs, Harris hip, UCLA, SF-12, and Western Ontario and McMaster Universities scores were obtained through 2 years. Median femoral head penetration was less among cross-linked compared to conventional liners as follows: 0.06 mm (0.04-0.08 mm) vs 0.08 mm (0.02-0.19 mm) at 6 months, 0.07 mm (−0.14 to 0.16 mm) vs 0.11 mm (0.01-0.27 mm) at 1 year, and 0.065 mm (−0.04 to 0.193 mm) vs 0.169 mm (0.09-0.22 mm) at 2 years. Clinical outcomes were similar between the groups. Highly cross-linked UHMWPE demonstrated 55% less femoral head penetration compared to conventional polyethylene at 2 years. Despite improvements in the manufacturing process and sterilization of conventional UHMWPE, the femoral head penetration rate is unchanged from historical standards.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , , ,