Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4063688 | The Journal of Arthroplasty | 2007 | 8 Pages |
Abstract
Audible squeaking in total hip replacements with ceramic-on-ceramic bearings is a rare problem. Acetabular component orientation was compared for 17 squeaking hips and 17 matched controls. Ninety-four percent of control hips were in a range of 25° ± 10° anteversion and 45° ± 10° inclination, but only 35% of squeaking hips were in this range (P = .0003). Eight hips squeak with bending. Four hips squeak with walking, and 5 hips squeak after prolonged periods of walking. Hips that squeaked with walking had acetabular components that were more anteverted (40°) than hips that squeaked with bending (19°) (P = .001) or prolonged walking (18°) (P = .020). The hips started squeaking after an average of 14 months. Patients with squeaking hips were younger, heavier, and taller than patients with silent hips.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
William L. MBBS, FRACS, FAOrthA, Gary C. FRCS (Tr & Orth), William K. MBBS, FRCS, FRACS, FAOrthA, Andrew MBBS, FRACS, FAOrthA, Bernard A. MD, FRACS, FAOrthA,