| Article ID | Journal | Published Year | Pages | File Type |
|---|---|---|---|---|
| 4063854 | The Journal of Arthroplasty | 2009 | 5 Pages |
Abstract
Duracon (Howmedica, Rutherford, NJ) posterior stabilized total knee system has a snap fit locking mechanism of a tibial polyethylene, including an additional locking screw for further fixation of polyethylene. We report 13 cases of locking screw migration from tibial component after Duracon posterior stabilized primary total knee arthroplasty. Among 13 knees, screw migration in 10 asymptomatic cases was incidentally detected during regular follow-up, and they were just observed in the outpatient clinic. Only 3 knees had moderate pain, swelling, and instability, and revision was done on 2 of 3 knees.
Keywords
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Woo-Shin MD, Yoon-Seok MD,