Article ID Journal Published Year Pages File Type
4063949 The Journal of Arthroplasty 2007 9 Pages PDF
Abstract

This in vitro biomechanical study compared a conventional balancing technique in knee arthroplasty to a technique using computer assistance. The experimental technique used a soft tissue tensioner instrumented with computer-monitored load cells to quantify soft tissue tension. To assess outcome, a tibial load transducer measured medial and lateral compartment forces and tibial rotation. An electromagnetic tracking system quantified knee position under simulated muscle loading. The computer-assisted technique improved knee balance before insertion of components. However, once components were implanted, there was no difference in knee load balance between the 2 techniques. No correlation was shown between compartmental load balance and tibial rotation or mechanical axis misalignment. Although computer-assisted surgery may improve technical accuracy, further work is necessary to achieve an optimal final knee load balance.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , , , ,