Article ID Journal Published Year Pages File Type
4063988 The Journal of Arthroplasty 2006 6 Pages PDF
Abstract

This study reviewed the long-term efficacy of the locking mechanism of the Harris-Galante porous (HGP) I and II acetabular components and the results and complications of polyethylene liner exchange without cement. There were 400 HGP-I components with a mean follow-up of 10 years (range, 2-19 years) and 78 HGP-II components with 8 years of mean follow-up (range, 2-13 years). There has been only 1 liner dislodgement (0.2%). Thirty-five hips (34 patients) have undergone liner exchange without cement. The index acetabular component was implanted as a primary procedure in 19 hips and a revision in 16 hips. No exchanged liner has dislodged at a mean follow-up time of 5.1 years (range, 2-11 years). However, there have been 7 patients (20%) with recurrent dislocation and all required reoperation. Dislocation was significantly lower when an elevated rim liner was used.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, ,