Article ID Journal Published Year Pages File Type
4064153 The Journal of Arthroplasty 2008 4 Pages PDF
Abstract

We report on 4 patients who underwent total knee arthroplasty with OtisKnee system (OtisMed, Hayward, Calif). An image-free computer navigation system was used to evaluate the deformities and the recommended cuts. The recommended custom cuts were as follows: valgus/varus cuts on the femur (5.5° valgus to 0.5° varus) in reference to the mechanical axis, flexion cuts on the femur (4°-9° of flexion); femoral rotation was within 1° of the epicondylar axis; valgus/varus cut on the tibia (3° of valgus to 7.5° of varus); tibial slope (5.5° of anterior slope to 0.5° of posterior slope). The custom OtisKnee system guides recommended alignment of the components that was more than 3° off of mechanical axis. The potential for malalignment with this system places implants at high risk of early failure.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
, , , ,