Article ID Journal Published Year Pages File Type
4076221 Journal of Shoulder and Elbow Surgery 2007 9 Pages PDF
Abstract

The objective of this study was to determine the effect of implant thickness and glenohumeral conformity on fixation of an all-metal glenoid component. A stainless steel glenoid component was designed and implanted in 10 cadaveric scapulae. A testing apparatus capable of producing a loading vector at various angles, magnitudes, and directions was used. The independent variables included 6 directions and 3 angles of joint load, 3 implant thicknesses, and 4 glenohumeral conformities. Implant micromotion relative to bone was measured by use of 4 displacement transducers at the superior, inferior, anterior, and posterior sites. The components displayed a consistent response to loading of ipsilateral compression and contralateral distraction. Stability decreased as the load application angle increased (P < .05). A decrease in the implant thickness and glenohumeral conformity resulted in increased implant stability (P < .05). Decreasing implant thickness and glenohumeral conformity reduce the eccentric component of loading and may improve the durability of glenoid implants.

Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation