Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4079334 | Operative Techniques in Orthopaedics | 2007 | 8 Pages |
Abstract
High tibial osteotomy (HTO) is a common therapy for medial gonarthrosis. Intraoperative planning and control of the correction is difficult. Therefore a workflow for a navigation with an image-free navigation system was developed for direct control of the correction of varus/valgus, internal/external rotation, and increase/decrease of the tibial slope. After registration of the lower limb, two k-wires define the osteotomy plane or planes for open- or closed-wedge osteotomy, respectively. The correction is accomplished under navigation system control in real-time. A rigid plate fixation is used for the fixation of the HTO. The reliability and accuracy of the system was approved with an aluminium phantom model.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
Stephan Lorenz, Mario Morgenstern, Andreas B. Imhoff,