Article ID Journal Published Year Pages File Type
4088348 Revue de Chirurgie Orthopédique et Réparatrice de l'Appareil Moteur 2008 4 Pages PDF
Abstract
Total hip arthroplasty was performed in November 2001 in a 30-year-old active patient with degenerative hip disease. A metal-on-metal bearing was used. Follow-up was considered excellent until the development of pain and squeaking at hip mobilization, leading to revision in March 2006. The acetabular metallic inlay of the metal-on-metal insert was found detached from the polyethylene insert; half of the diameter of the neck of the femoral stem was sectioned. Complete revision was performed with an acetabular graft. At one year follow-up, anatomic and functional outcome has been excellent. To our knowledge, this is the first report of this kind of mechanical failure of a metal-on-metal total hip arthroplasty.
Related Topics
Health Sciences Medicine and Dentistry Orthopedics, Sports Medicine and Rehabilitation
Authors
,