Abstract
Alumina ceramic-on-ceramic bearings have very low wear rates in standard hip simulators. However retrieved bearings demonstrate edge loading wear. We analyzed 16 retrieved bearings. The average wear volume was 0.4 mm3/year in the femoral heads. Acetabular anteversion of less than 15° was associated with an increased wear rate. Rarely patients complain of squeaking from their hip replacement. Of hip arthroplasties that squeak, those that squeak in extension have more anteversion (40°) than those that squeaked in flexion (19°). The incidence of both stripe wear and squeaking will be reduced by accurate placement of the hip components.
Related Topics
Health Sciences
Medicine and Dentistry
Orthopedics, Sports Medicine and Rehabilitation
Authors
W.L. Walter, P.J. Lusty, A. Watson, G. O’Toole, M.A. Tuke, B. Zicat, W.K. Walter,