Article ID Journal Published Year Pages File Type
4125540 Otolaryngology - Head and Neck Surgery 2009 4 Pages PDF
Abstract
To provide assistance to otolaryngologists to decide the best manner in which to reprocess flexible nasopharyngoscopes, a review of existing English language medical literature regarding the methods of flexible endoscope reprocessing was performed, including previously published guidelines from other medical disciplines. Multiple steps were confirmed to be critical to effectively reprocess flexible nasopharyngoscopes. High-level disinfection has been determined to be the minimum level of disinfection required for reprocessing of flexible nasopharyngoscopes. Several steps are important in all reprocessing techniques, including manual cleaning, leak testing, cleaning with an enzymatic agent, high-level disinfection, and drying with vertical storage. Three techniques are available to achieve high-level disinfection: manual disinfection with a liquid disinfectant/sterilant, use of an automated endoscope reprocessor, and use of a disposable sheath. Achieving high-level disinfection of flexible nasopharyngoscopes can be accomplished by a variety of methods. Strict adherence to recommended procedures is critical.
Related Topics
Health Sciences Medicine and Dentistry Otorhinolaryngology and Facial Plastic Surgery
Authors
,