Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4204699 | Archivos de Bronconeumología | 2007 | 4 Pages |
Abstract
Measurement of the fraction of exhaled nitric oxide (FENO) provides a noninvasive way to monitor asthma treatment in clinical practice. The aim of this study was to determine FENO reference values for measurements recorded with the portable NIOX MINO monitor in a group of healthy volunteers. We also assessed the association between values recorded by the portable monitor and the N-6008 chemoluminescence analyzer used in our pulmonary function laboratory. The FENO values obtained with the portable monitor were consistently higher than those recorded by the N-6008 analyzer; the cutoff value for the portable monitor was 34 ppb (mean + 2 standard deviation). We detected a direct correlation (r=0.92) between the FENO measurements recorded by the 2 monitors (P=.001). The following equation expresses the relationship between measurements from the 2 devices: FENO(NIOX MINO) = 10 + [1.5 FENO(N-6008)]. We did not observe statistically significant correlations between FENO measurements and age, sex, body mass index, or spirometry.
Related Topics
Health Sciences
Medicine and Dentistry
Pulmonary and Respiratory Medicine
Authors
Ana MarÃa Fortuna, Teresa Feixas, Pere Casan,