No catheter failures were demonstrated in this study using injection rates well above those used in conventional CTA. Power injector tubing failure was demonstrated at an injection rate of 30 ml/sec, which generated mean pressures in the 16-cm catheters of 920 psi (tubing rating per manufacturer is 300 psi). This study demonstrated no catheter or injector tubing failure at injection rates of 5 to 25 ml/sec.